Activity-Specific Tremor Quantification Using Wearable Accelerometer and Gyroscope Reveals Differential Treatment Response Following Focused Ultrasound Thalamotomy in Essential Tremor
Objective: Clinical assessment of tremor following MR-guided focused ultrasound (MRgFUS) thalamotomy relies on subjective rating scales such as the Clinical Rating Scale for Tremor, which provide limited insight into activity-specific treatment response and cannot distinguish translational from rotational tremor components. Wearable inertial measurement units (IMUs) may enable objective, multi-dimensional tremor characterization across motor tasks. We […]
See abstract details
Objective:
Clinical assessment of tremor following MR-guided focused ultrasound (MRgFUS) thalamotomy relies on subjective rating scales such as the Clinical Rating Scale for Tremor, which provide limited insight into activity-specific treatment response and cannot distinguish translational from rotational tremor components. Wearable inertial measurement units (IMUs) may enable objective, multi-dimensional tremor characterization across motor tasks. We aimed to quantify activity-specific and modality-specific tremor changes following MRgFUS thalamotomy using a wearable 6-axis IMU in essential tremor (ET) patients.
Methods:
Ten ET patients undergoing MRgFUS thalamotomy at North Shore University Hospital (Manhasset, NY) performed four standardized motor tasks (rest, arm extension, arm flexion at 90 degrees, and a functional drinking task), each for one minute, immediately before and after the procedure. A wrist-worn Shimmer3 IMU recorded tri-axial acceleration and angular velocity at 512 Hz. Tremor was quantified as spectral power in the 4–12 Hz band from 5-second windowed segments per task. Per-patient percentage change was summarized across the cohort.
Results:
Tremor power decreased consistently during postural and functional tasks, with all 10 patients showing improvement during arm flexion (accelerometer: median reduction 88.1%, range 25.8–99.9%) and the drinking task (median 89.4%, range 25.0–99.8%). Nine of 10 patients with measurable extension tremor improved post-lesion (accelerometer: median reduction 91.0%); the remaining patient had no baseline extension tremor. Rest tremor metrics were highly variable, consistent with ET postural-kinetic phenotype. Gyroscope-derived tremor power showed the most consistent treatment response during the functional drinking task (-85.1% ± 14.3% vs. -80.3% ± 22.4% for accelerometer), suggesting rotational tremor may be particularly responsive to thalamic lesioning. The dominant tremor frequency (~5–6 Hz) was preserved post-treatment despite substantial amplitude reduction (peak spectral power reduced by 83–86% across arm flexion and drinking tasks), indicating oscillatory amplitude suppression without altering the underlying tremor frequency.
Conclusions:
Wearable 6-axis IMU provides objective, activity-specific tremor quantification that reveals differential treatment responses across tasks and sensor modalities not captured by conventional clinical scales, complementing patient assessments. The functional drinking task yielded the most consistent metrics across both sensor modalities, offering a novel way to track tremor improvement that correlates with quality of life.